Induced atrial fibrillation during defibrillation test on implantation of subcutaneous implantable cardioverter defibrillator
Ki-Hun Kim
Sang-Hoon Seol
Jino Park
Yeo-Jeong Song
Seunghwan Kim
Dong-Kie Kim
1.Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea2.Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea3.Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea4.Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea5.Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea6.Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine, Busan, Korea
摘要:The subcutaneous implantable cardioverter defibrillator (S-ICD) has become an alternative to the transvenous ICD in indicated patients.However, inappropriate shock or failed ventricular tachycardia/fibrillation conversion is the most alarming complication of S-ICD.Therefore, defibrillation test (DFT) is recommended for the S-ICD implantation.However, the DFT can provoke or worsen cardiac arrhythmias.We present two induced atrial fibrillation (AF) cases by DC bursts on DFT during S-ICD implantation.The AF was not terminated by the 65-J S-ICD shock, even when ventricular fibrillation was terminated.