Lead dislodgement complicated with paroxysmal atrial fibrillation as a rare cause of inappropriate implantable cardioverter defibrillator shock:a case report
Gao-Feng LIN1
Tong ZOU1
Ming-Hui DU2
Jia-Bin TONG2
Hai-Feng SHI2
Hao CHEN2
Jun-Peng LIU2
Jia CHONG2
You LYU2
Su-Juan WU2
Zhi-Lei WANG2
Xin JIN2
Xu GAO2
Lin SUN2
Jie-Fu YANG1
1.Department of Cardiology,Beijing Hospital,National Center of Gerontology,Institute of Geriatric Medicine,Chinese Academy of Medical Sciences,Beijing,China;Graduate School,Peking Union Medical College,Chinese Academy of Medical Sciences,Beijing,China2.Department of Cardiology,Beijing Hospital,National Center of Gerontology,Institute of Geriatric Medicine,Chinese Academy of Medical Sciences,Beijing,China
摘要:Inappropriate shock of implantable cardi-overter defibrillator(ICD)is a rare but cata-strophic phenomenon.Inappropriate shock is usually caused by oversensing of noise or supra-ventricular tachycardias(SVTs).However,inappro-priate shock can also be triggered by other uncom-mon mechanisms.Lead dislodgement is also rare in patients with cardiac implantable electronic devices,which can manifest as pacing failure and mis-sensing of other signals.Here,we report a rare case of inap-propriate ICD shock triggered by lead dislodge-ment complicated with paroxysmal atrial fibrilla-tion(PAF)in an elderly patient.