Erythrocyte Fragment Count Predicts Hemolysis in Roller Pumps
FAN Jun-qiang
XU Shi-wei
CHEN Fang
DING Min-jun
摘要:Hemolysis in blood pumps has been measured by various in vitro test methods, in which normalized index of hemolysis (NIH) was established. As NIH is complicated and difficult to calculate, erythrocyte fragment count is proposed in the present study to predict hemolysis in roller pumps. Methods: Five paired in vitro tests wereconducted using the POLYSTAN pediatric pump(group A) and COBE pump(group B). Ten whole blood samples (400 ml ) were circled in the roller pump for 16 h. Erythrocyte fragments count and plasma-free hemoglobin (FHb) were measured before pumping and every two hours through circulation after four-hour-pumping. The morphological changes of erythrocyte were observed by scanning electron microscope. Results: The two groups' EFC and FHb levels were increased linearly during a long duration of pumping and linear regression of erythrocyte fragments count and plasma-free hemoglobin were correlated. Conclusion: Erythrocyte fragments count could be used as an index in evaluating the in vitro hemolytic properties of blood pumps.
机标关键词:scanning electron microscopein vitroblood pumpsmorphological changeslinear regressiontest methodswhole blood
分类号:R318.11(医用一般科学)
论文发表日期:2007-01-01
在线出版日期:2025-08-15(本平台首次上网日期,不代表文献的发表时间)
页数:6( 33-38 )