Magnetic resonance imaging after cochlear implants
Naif Bawazeer
Hella Vuong
Sophie Riehm
Francis Veillon
Anne Charpiot
摘要:Introduction:Most cochlear implants are currently compatible with magnetic resonance imaging (MRJ) up to 3 T.Nevertheless,this does not completely eliminate the risk of serious accidents.Implant displacements and other adverse events with compatible implants have been reported in the literature.Case reports:Among the six patients who had MRI after receiving implants at our center,we report three cases with adverse events related to the examination.The first case was complicated by magnet displacement with partial demagnetization.The second case showed total demagnetization,which necessitated removal and reimplantation of the implant.The third case involved severe pain sensation which disrupted the MRI scan.The smallest artifact was found with 3D MRI angiography,and largest artifact was found with diffusion and T2 FLASH.Discussion:Moving the patient into the MRI apparatus must be supervised by an otorhinolaryngology specialist or an experienced radiologist.It is important to consider the magnetic field directions,so that angle between the implant magnetic fields and the MRI BO always remains less than or equal to 90°.In addition,we recommend the use of an “arrow drawing” to facilitate the orientation of the magnetic field directions.Furthermore,to prevent magnet displacement,we recommend systematic use of a protective splint in addition to bandaging.