Calorimeter stability test and effect of ignition heat fixed value calibration on calorific value
GONG Fu
Abstract:By tracking the calorific capacity stability and testing the calorific value of standard material, the effect of ignition heat calibration methods on determination result was analyzed to determine the accuracy of the instrument and control method.The track test results showed that calorimeter stability was good, the accuracy of determination result of benzoic acid and standard material meet the requirements, there was no significant difference between different instruments.Comparison between fixed value calibration and actual consumption calibration of ignition heat showed that the error through fixed value calibration can be ignored.
Keywords:calorimetercalorific capacitystabilityignition heat calibrationcalorific value
Publication Date:2016-01-01
Pages:6( 42-46,70 )

ISSN:1007-7677
Year, Vol.(Issue):2016,(6)